Generator compatible is a sentence, not a specification. A theatre that dips when the sterilizers start needs three documents before any purchase order is credible: a site power condition sheet filled in by the hospital with measured values, a per-model ride-through declaration from the factory that states what the table and the light do at those values, and an on-site dip test at commissioning that records the equipment’s behaviour and gets signed. The site sheet comes first. Until the hospital knows its own residual voltage, duration, recurrence and generator load, it cannot judge any factory claim, and the factory cannot be asked to meet a condition nobody has written down.
Why the Compliance Statement Does Not Answer the Question
Immunity testing and site compatibility are different claims. A unit that passes the electromagnetic immunity tests in IEC 60601-1-2, including the voltage-dip and short-interruption tests of IEC 61000-4-11, has demonstrated that it meets the standard’s performance criteria at the standard’s test levels on a laboratory source. That is a laboratory statement about a defined stimulus. It says nothing about what happens on a hospital generator when a sterilizer bank draws starting current, because that event has its own depth, duration, repetition and source impedance.
The two claims can be separated cleanly by asking for the test conditions. What voltage was the source? What was the residual voltage as a percentage of nominal, and for how many cycles? Which performance criterion was reached, A, B or C? Was the unit powered from a generator or from a programmable AC source? Was the equipment under load, and were its motors or its battery being charged at the time? A declaration that answers those questions is a specification. One that says compliant does not.
The criteria deserve a sentence each, because the difference decides clinical acceptance. Criterion A means the equipment behaved normally. Criterion B means a temporary degradation that corrected itself without operator action. Criterion C means the operator had to intervene, reset or restart. In a theatre, criterion C during an operation is the case the whole document set exists to prevent.
Fill In the Site Sheet Before You Ask Any Factory
The sheet has six fields, and every one of them is a measurement rather than an opinion. Nominal voltage and frequency. Residual voltage during a real event, as a percentage of nominal. Duration of the dip in milliseconds or cycles. Recurrence, meaning how many events occur within a defined period and how closely they cluster. Generator load and source condition at the time of the measurement. And the point in the distribution where the measurement was taken.
The measurement method matters as much as the numbers. A power quality recorder at the theatre’s distribution board captures the event as the theatre sees it. A logger at the equipment isolator captures it closer to the load. Both are useful and they are not interchangeable, because voltage drop along the feeder is part of the problem. Record at the board if the question is what the building delivers, and at the isolator if the question is what the equipment receives.
Timing is the next trap. A single weekday morning sample will miss the event that matters, because the dip is caused by a specific load switching on. Measure across a full working day that includes the sterilizer cycles, and annotate the recording with what was running. If the hospital has never measured, the honest entry on the sheet is not measured, and that entry is itself a finding: the requirement is being specified from assumption.
What the Ride-Through Declaration Must State
Ask for one page per model, in a fixed format, so two suppliers can be compared. The page should state the supply envelope the unit is designed for, the hold-up time of its internal supply, and then the behaviour at a series of defined events, each with the event written as percentage of nominal voltage and duration in cycles.
| Field | What a usable answer contains | Why it matters on site |
|---|---|---|
| Declared supply envelope | Voltage and frequency range, phases, steady state tolerance, maximum inrush | Confirms whether the site’s measured dips sit inside the declaration or outside it |
| Hold-up time | The duration the internal supply maintains its output after input is lost | Predicts whether a dip shorter than the hold-up time is invisible to the user |
| Event table | Behaviour at each stated dip and interruption, with the criterion reached | Gives the commissioning test something to check against |
| Battery and emergency behaviour | What the battery can operate, for how long, and what is unavailable on battery | Decides whether a table can still be moved and a light still gives useful field |
| Recovery | Automatic recovery or operator reset, time to normal output, settings retained or lost | Separates criterion B from criterion C in clinical terms |
| Test provenance | Generator or programmable source, load condition, software version, report reference | Shows whether the declaration came from the factory’s own test or from a datasheet assumption |
The Table and the Light Answer Differently
A table’s power behaviour is a safety question about position. The relevant product standard is IEC 60601-2-46, the particular requirements for operating tables, and the point buyers should carry into the tender is that an interruption of the supply must not cause the table to change its configuration. Nothing should drop, nothing should drift, no brake should release. The declaration should also say which functions remain available on battery or emergency operation and which do not, because a table that cannot be raised during a power failure is a different clinical proposition from one that can.
A light’s power behaviour is a question about illumination over a patient. The relevant standard is IEC 60601-2-41, the particular requirements for surgical luminaires and luminaires for diagnosis, which sets recovery requirements after an interruption rather than treating the light as a constant. The tender question is therefore not does the light stay on, but what the measured illuminance is at one second, at five seconds and at one minute after a dip, and whether the light restores its previous intensity and field by itself or needs a hand.
Two other behaviours belong in the same answer because they are the ones staff notice. Does the light flash or change colour temperature during the dip, and does the field size or focus shift when output comes back? A light that recovers its centre illuminance but has moved its pattern is a light that has to be repositioned mid case.
The On-Site Dip Test at Commissioning
The commissioning test is the step that turns the declaration into evidence for this building. Run it once per unit, at the equipment isolator, with the unit configured as it will be used. If the site has a load bank, use it to create a repeatable event. If it does not, use the actual cause: start the sterilizer bank and record the dip the theatre sees while the table and the light are live.
Record the whole event, not the aftermath. The trace should show pre-event voltage, the residual voltage, the duration, the recovery curve and the time to normal output. Note the generator state and the load at the moment of the event, the software and configuration of the equipment, the battery state of charge, and which functions were being operated. Then write the observed behaviour against the declaration line by line and sign it with the biomedical engineer and the factory’s commissioning engineer.
Three rules keep the test honest. The unit must be tested as installed, because a factory bench test does not reproduce the site’s earthing, feeder or generator. The test must be repeated after any change to the supply path, including a new sterilizer, a generator service or a switchgear change. And the trace must be filed with the room record, not with the supplier’s project file, so the next periodic test can compare like with like.
Pendant, Shared Feeder and Downstream Loads
Most theatre power arguments are arguments about topology. The pendant, the table and the light may share a feeder. The sterilizer may sit downstream of the same board. The generator may serve the theatre and the laundry. Two rooms fitted with identical equipment can therefore behave differently at the same hospital, and neither the factory nor the electrical contractor is wrong on paper.
Draw the topology into the site sheet: which board feeds which room, what else is on that board, where the sterilizer connects, where the generator transfer switch sits and what the generator also serves. Then measure at the point where each piece of equipment is connected. If the equipment shares a feeder with the sterilizer, the commissioning test must recreate that shared condition rather than testing on a clean source.
The pendant deserves its own line because it is often the first thing to misbehave. Its brakes, its lighting and any integrated services draw from the same supply as the room, and its behaviour during a dip should be declared and observed in the same test as the table and the light. The electrical variant of the equipment also has to be frozen before this work starts, since a unit built for a different voltage or plug configuration cannot be tested against the site sheet: the discipline is set out in freezing the electrical variant.
Year Two: Battery Capacity and the Periodic Re-Test
The dip behaviour declared at commissioning decays with the batteries. A light that held its field through a five second interruption when new may hold it for less time after two years of charging cycles, and the change is invisible until the day it matters. Put the measured runtime into the acceptance record and then measure it again on a schedule.
The practical form is a capacity test under load, recorded as a runtime figure, with the date codes of the batteries noted at the same time. The commissioning and battery discipline that supports this is described in commissioning the battery before you judge it, and the specification side, including charging behaviour and replacement intervals, is covered in specifying batteries and charging. Whenever the runtime drops below the declared figure, the unit’s dip behaviour is no longer the declared behaviour, and that is a maintenance finding rather than a surprise.
Fold the dip check into the preventive maintenance visit as well. The visit can be short: start the sterilizer or the load bank, record the event, compare the recovery with the baseline trace, and note the battery runtime. The manufacturer’s preventive maintenance scope should be written to include it, which is the same document set discussed in what to ask for in a preventive maintenance schedule.
Clause Wording for the Order
Four sentences do most of the work. The supplier shall state the site conditions assumed for correct operation and shall declare ride-through behaviour per model at the site’s measured conditions. The supplier shall provide the dip and interruption behaviour in the format of the event table, with the performance criterion reached at each event. Acceptance shall include an on-site dip test at commissioning, recorded as a trace and signed by both parties, with the observed behaviour matched against the declaration. The warranty shall cover any unit whose measured behaviour at commissioning departs from its declaration.
The order of work is what makes the set effective. Measure the site first. Put the measured values into the tender. Require the declaration against those values. Test at commissioning and file the trace. Re-measure when the batteries age or the distribution changes. A factory that can answer all of that is a factory that has tested its own equipment, and a hospital that can ask for it is a hospital that will not discover its generator problem during an operation.