{"id":4178,"date":"2026-09-15T11:01:18","date_gmt":"2026-09-15T11:01:18","guid":{"rendered":"https:\/\/sanyangmedical.com\/medical-trolley-load-capacity-testing\/"},"modified":"2026-09-21T17:17:51","modified_gmt":"2026-09-21T17:17:51","slug":"medical-trolley-load-capacity-testing","status":"publish","type":"post","link":"https:\/\/sanyangmedical.com\/id\/medical-trolley-load-capacity-testing\/","title":{"rendered":"Medical Trolley Load Capacity: How Static and Dynamic Ratings Are Tested"},"content":{"rendered":"<p><!-- egp:no-toc --><\/p>\n<p class=\"quick-answer\"><strong>Quick answer:<\/strong> A medical trolley sold as a 60 kg static rating is usually safe to 40 kg dynamic, because rolling and braking loads add stress that a parked test does not capture. Buyers who order on a single headline figure end up with carts that fail at the castor rather than the shelf.<\/p>\n<h2>What Static And Dynamic Ratings Mean<\/h2>\n<p>Two numbers describe the same cart. Static capacity is the load a frame carries while parked, and dynamic capacity is the load it carries while moving, cornering and braking. A 60 kg static cart holds about 40 kg in motion, and the difference sits in the castors, the brake pads and the frame joints rather than in the shelf surface.<\/p>\n<ul>\n<li>Static rating: measured with the cart parked on a level floor.<\/li>\n<li>Dynamic rating: measured while rolling over a 10 mm threshold.<\/li>\n<li>Brake hold: measured on a slope with the full rated load.<\/li>\n<li>Drawer rating: measured per drawer, not for the whole cart.<\/li>\n<\/ul>\n<h2>How A Load Test Is Run In The Factory<\/h2>\n<p>A trolley test sheet is short and repeatable. Weight plates are loaded evenly, the cart rolls a set distance, and the frame is measured before and after. The sequence below reflects powder-coated steel carts with 100 mm castors.<\/p>\n<table>\n<tr>\n<th>Test<\/th>\n<th>Load applied<\/th>\n<th>Duration<\/th>\n<th>Pass criteria<\/th>\n<th>Evidence kept<\/th>\n<\/tr>\n<tr>\n<td>Static top load<\/td>\n<td>90 kg (1.5 times rated)<\/td>\n<td>24 hours<\/td>\n<td>deflection under 2 mm<\/td>\n<td>photo plus test sheet<\/td>\n<\/tr>\n<tr>\n<td>Dynamic rolling<\/td>\n<td>60 kg rated load<\/td>\n<td>2 hours<\/td>\n<td>castor wear under 1 mm<\/td>\n<td>before and after photos<\/td>\n<\/tr>\n<tr>\n<td>Brake hold on 5 degrees<\/td>\n<td>60 kg rated load<\/td>\n<td>10 minutes<\/td>\n<td>movement under 5 mm<\/td>\n<td>video clip plus sheet<\/td>\n<\/tr>\n<tr>\n<td>Drawer cycling<\/td>\n<td>20 kg per drawer<\/td>\n<td>5,000 times<\/td>\n<td>runner play under 3 mm<\/td>\n<td>cycle counter readout<\/td>\n<\/tr>\n<tr>\n<td>Packed carton drop<\/td>\n<td>carton weight, 800 mm height<\/td>\n<td>1 drop per corner<\/td>\n<td>frame alignment within 2 mm<\/td>\n<td>carton condition photos<\/td>\n<\/tr>\n<\/table>\n<figure><img decoding=\"async\" src=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/load-test-trolley.png\" alt=\"Medical trolley load test with weight plates on a steel cart\"><figcaption>Weight plates plus a dial indicator answer the capacity question in one day.<\/figcaption><\/figure>\n<h2>Castors, Brakes And The 100 mm Rule<\/h2>\n<p>Castor diameter sets rolling effort. A 100 mm castor rolls a 60 kg load with noticeably less push force than a 75 mm castor, and hospital staff feel the difference within a week of use. Two castors should carry brakes on a four-castor cart, and the brake should be a total lock rather than a directional lock if the cart parks on a ramp.<\/p>\n<p>Wheel material matters as much as size. Polyurethane tyres protect vinyl flooring and resist flat spots, while hard nylon tyres roll easier but mark soft floors. A hospital that mixes both materials along one corridor will replace castors more often than frames, and castor replacement sits in the maintenance record rather than the purchasing file.<\/p>\n<h2>Specifying Capacity For A Tender<\/h2>\n<p>Tender documents should state four values separately: static capacity in kg, dynamic capacity in kg, per-drawer capacity in kg and castor diameter in mm. A cart quoted against a single figure invites substitution, and a substituted castor can cut the dynamic rating by 30 percent without changing the frame at all.<\/p>\n<p>Two controls keep the specification alive. Ask for the test sheet with the batch number on it, and require the packed carton weight on the packing list so a lighter build becomes visible before delivery. On multi-cart orders, re-test the first 10 units of each production run against the approved sample and photograph the castor mounting plate, which is where most substitutions appear.<\/p>\n<p>A receiving check closes the loop. Weigh the first carton of every delivery against the packing list, roll one cart over a 10 mm threshold with the rated load, and compare the castor plate against the approved sample photo. The check takes 20 minutes per delivery and catches a substitution before the carts reach the wards.<\/p>\n<figure><img loading=\"lazy\" decoding=\"async\" src=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/ward-quiet-trolley.png\" alt=\"Ward quiet trolley with low noise castors for hospital use\"><figcaption>Low noise castors change the ward noise profile long before the frame wears.<\/figcaption><\/figure>\n<div class=\"yt-facade\" data-yt-id=\"s13FiO0Goiw\" role=\"button\" tabindex=\"0\" aria-label=\"Watch: medical equipment manufacturing overview\"><span class=\"yt-facade-label\">&#9654; Watch: medical equipment manufacturing overview<\/span><\/div>\n<h2>Frequently Asked Questions<\/h2>\n<div class=\"faq-zone\">\n<h3>Is static or dynamic load the number to buy on?<\/h3>\n<p>Dynamic. A cart pushed between departments never performs at its static figure, and 40 kg against 60 kg static is normal for steel carts with 100 mm castors.<\/p>\n<h3>How long should a brake hold test run?<\/h3>\n<p>Ten minutes on a 5 degree slope with the rated load and movement under 5 mm. Shorter tests miss the slow creep that reaches maintenance teams first.<\/p>\n<h3>What castor diameter suits a loaded hospital cart?<\/h3>\n<p>100 mm for carts up to 60 kg dynamic, and 125 mm above that. Smaller castors raise push force sharply on floors with joints or a threshold.<\/p>\n<h3>What evidence should ship with a trolley order?<\/h3>\n<p>The load test sheet with batch number, carton and packing list weights, and photos of the castor mounting plate. Those three make a substitution visible.<\/p>\n<\/div>\n<p>Buyers comparing frames before a tender can review the range on <a href=\"https:\/\/sanyangmedical.com\/medical-trolleys\/\">medical trolleys<\/a> and the drawer and accessory choices on <a href=\"https:\/\/sanyangmedical.com\/medical-trolley-accessories-configurations\/\">trolley accessories and configurations<\/a>. Brake and lock hardware maintenance sits in the note on <a href=\"https:\/\/sanyangmedical.com\/trolley-lock-cylinder-maintenance-keyed-push-lock\/\">trolley lock cylinder maintenance<\/a>. Independent test methods for medical carts follow the inspection practice published by <a href=\"https:\/\/www.sgs.com\/\" target=\"_blank\" rel=\"noopener\">SGS<\/a>.<\/p>\n<p><script type=\"application\/ld+json\">\n{\"@context\":\"https:\/\/schema.org\",\"@type\":\"FAQPage\",\"mainEntity\":[\n{\"@type\":\"Question\",\"name\":\"Is static or dynamic load the number to buy on?\",\"acceptedAnswer\":{\"@type\":\"Answer\",\"text\":\"Dynamic. A cart pushed between departments never performs at its static figure, and 40 kg against 60 kg static is normal for steel carts with 100 mm castors.\"}},\n{\"@type\":\"Question\",\"name\":\"How long should a brake hold test run?\",\"acceptedAnswer\":{\"@type\":\"Answer\",\"text\":\"Ten minutes on a 5 degree slope with the rated load and movement under 5 mm. Shorter tests miss the slow creep that reaches maintenance teams first.\"}},\n{\"@type\":\"Question\",\"name\":\"What castor diameter suits a loaded hospital cart?\",\"acceptedAnswer\":{\"@type\":\"Answer\",\"text\":\"100 mm for carts up to 60 kg dynamic, and 125 mm above that. Smaller castors raise push force sharply on floors with joints or a threshold.\"}},\n{\"@type\":\"Question\",\"name\":\"What evidence should ship with a trolley order?\",\"acceptedAnswer\":{\"@type\":\"Answer\",\"text\":\"The load test sheet with batch number, carton and packing list weights, and photos of the castor mounting plate. Those three make a substitution visible.\"}}]}\n<\/script><br \/>\n<script type=\"application\/ld+json\">\n{\"@context\":\"https:\/\/schema.org\",\"@type\":\"Article\",\"headline\":\"Medical Trolley Load Capacity: How Static and Dynamic Ratings Are Tested\",\"description\":\"Medical trolley load capacity explained through the factory test sequence: static and dynamic ratings, brake hold on a slope, drawer cycling and castor diameter rules.\",\"about\":[\"medical trolley load capacity\",\"dynamic load rating\",\"castor selection\"],\"keywords\":\"medical trolley load capacity, hospital cart load test, castor rating\",\"articleSection\":\"Surgical Equipment Guides\",\"data_statements\":[\"A 60 kg static cart usually carries 40 kg dynamic\",\"Static test applies 90 kg for 24 hours with deflection under 2 mm\",\"Brake hold runs 10 minutes on a 5 degree slope with movement under 5 mm\",\"Drawer cycling runs 5,000 times at 20 kg per drawer\",\"100 mm castors suit carts up to 60 kg dynamic, 125 mm above that\"],\"qa_concise\":\"Medical trolley load capacity should be bought on the dynamic figure, which runs about two thirds of the static rating once rolling and braking loads are counted.\"}\n<\/script> See also our <a href=\"https:\/\/sanyangmedical.com\/castor-selection-hospital-trolleys-floors-braking\/\">hospital trolley castor guide<\/a>.<\/p>\n<p><strong>Related reading:<\/strong> <a href=\"https:\/\/sanyangmedical.com\/medical-suction-trolley-selection\/\">medical trolley load capacity testing<\/a>.<\/p>\n","protected":false},"excerpt":{"rendered":"<p>Quick answer: A medical trolley sold as a 60 kg static rating is usually safe to 40 kg dynamic, because rolling and braking loads add stress that a parked test does not capture. Buyers who order on a single headline figure end up with carts that fail at the castor rather than the shelf. What &#8230; <a title=\"Medical Trolley Load Capacity: How Static and Dynamic Ratings Are Tested\" class=\"read-more\" href=\"https:\/\/sanyangmedical.com\/id\/medical-trolley-load-capacity-testing\/\" aria-label=\"Baca selengkapnya tentang Medical Trolley Load Capacity: How Static and Dynamic Ratings Are Tested\">Baca Selengkapnya<\/a><\/p>","protected":false},"author":1,"featured_media":3253,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"rank_math_title":"Medical Trolley Load Capacity: Static vs Dynamic Ratings","rank_math_description":"Medical trolley load capacity tested on the factory floor: 90 kg static against 60 kg dynamic, brake hold on a 5 degree slope and the castor rule for hospital carts.","rank_math_focus_keyword":"medical trolley load capacity","rank_math_robots":"","rank_math_canonical_url":"","rank_math_facebook_title":"","rank_math_facebook_description":"","rank_math_twitter_title":"","rank_math_twitter_description":"","_yoast_wpseo_title":"","_yoast_wpseo_metadesc":"","_yoast_wpseo_focuskw":"","_yoast_wpseo_canonical":"","_yoast_wpseo_meta-robots-noindex":"","_yoast_wpseo_meta-robots-nofollow":"","_yoast_wpseo_opengraph-title":"","_yoast_wpseo_opengraph-description":"","_yoast_wpseo_twitter-title":"","_yoast_wpseo_twitter-description":"","_aioseo_title":"","_aioseo_description":"","_aioseo_keywords":"","_aioseo_robots_default":"","_aioseo_robots_noindex":"","_aioseo_og_title":"","_aioseo_og_description":"","_aioseo_twitter_title":"","_aioseo_twitter_description":"","aiosp_title":"","aiosp_description":"","aiosp_keywords":"","_seopress_titles_title":"","_seopress_titles_desc":"","_seopress_analysis_target_kw":"","_seopress_robots_canonical":"","_seopress_robots_index":"","_seopress_robots_follow":"","_seopress_social_fb_title":"","_seopress_social_fb_desc":"","_seopress_social_twitter_title":"","_seopress_social_twitter_desc":"","_genesis_title":"","_genesis_description":"","_genesis_canonical":"","_genesis_noindex":"","_genesis_nofollow":"","slim_seo":"","footnotes":""},"categories":[108],"tags":[],"class_list":["post-4178","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-surgical-equipment-guides"],"_links":{"self":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/4178","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/comments?post=4178"}],"version-history":[{"count":2,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/4178\/revisions"}],"predecessor-version":[{"id":4252,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/4178\/revisions\/4252"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/media\/3253"}],"wp:attachment":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/media?parent=4178"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/categories?post=4178"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/tags?post=4178"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}