{"id":3734,"date":"2026-08-12T11:49:19","date_gmt":"2026-08-12T11:49:19","guid":{"rendered":"https:\/\/sanyangmedical.com\/operating-table-manual-override-emergency-crank\/"},"modified":"2026-08-20T21:10:02","modified_gmt":"2026-08-20T21:10:02","slug":"operating-table-manual-override-emergency-crank","status":"publish","type":"post","link":"https:\/\/sanyangmedical.com\/id\/operating-table-manual-override-emergency-crank\/","title":{"rendered":"Operating Table Manual Override: Emergency Crank Operation Explained"},"content":{"rendered":"<p><!-- Rank Math SEO --><br \/>\n<!-- Title: Operating Table Manual Override: Emergency Crank Operation | Sanyang --><br \/>\n<!-- Slug: operating-table-manual-override-emergency-crank --><br \/>\n<!-- Focus Keyword: operating table manual override --><br \/>\n<!-- Meta Description: Operating table manual override explained: how the emergency crank and hydraulic release valve work, what IEC 60601-2-46 requires, and what buyers must test before purchase. --><br \/>\n<!-- Category: surgical-equipment-guides --><\/p>\n<p>When power fails mid-procedure or a drive motor jams, the operating table manual override is the difference between a controlled repositioning and a stalled surgery. Every electric and electro-hydraulic surgical table carries a mechanical fallback \u2014 usually an emergency crank or a manual release valve \u2014 that lets staff move the tabletop without electricity. This article explains where those controls sit, how they work, and what a buyer should verify before accepting a table into an operating room.<\/p>\n<p>If the crank feels spongy or you spot oil around the base column, stop and inspect for seal failure first: our <a href=\"https:\/\/sanyangmedical.com\/operating-table-hydraulic-oil-leak-repair-2\/\">operating table hydraulic oil leak repair guide<\/a> walks through the $150 in-house fix.<\/p>\n<p>The override gets little attention during procurement. Spec sheets list it as a line item, sales teams mention it once, and the crank sits in its bracket for years. Then a utility outage or actuator failure happens, and the team that never touched the crank has to use it under pressure. Understanding the override before purchase, not after an incident, is the point of this guide.<\/p>\n<div class=\"yt-facade\" data-yt-id=\"W63vrbX-mLg\" role=\"button\" tabindex=\"0\" aria-label=\"Play video: Safe Patient Transfer Procedure: Surgical Teaching Unit - McGill University - JGH\" style=\"position:relative;width:100%;padding-bottom:56.25%;background:#000;border-radius:8px;margin:24px 0;cursor:pointer;overflow:hidden;\">\n<div style=\"position:absolute;top:0;left:0;width:100%;height:100%;display:flex;align-items:center;justify-content:center;\">\n<svg width=\"68\" height=\"48\" viewBox=\"0 0 68 48\"><path d=\"M66.52 7.5c.78 2.94.78 15 0 17.94a8.5 8.5 0 01-6 6C57.58 32.22 34 32.22 34 32.22s-23.58 0-26.52-.78a8.5 8.5 0 01-6-6C.7 22.5.7 10.44 1.48 7.5a8.5 8.5 0 016-6C10.42.72 34 .72 34 .72s23.58 0 26.52.78a8.5 8.5 0 016 6z\" fill=\"red\"\/><path d=\"M45 24L27 14v20\" fill=\"white\"\/><\/svg>\n<\/div>\n<\/div>\n<h2>What Manual Override Means on an Operating Table<\/h2>\n<p>Manual override is the set of mechanical controls that move or lower the tabletop when the primary drive \u2014 electric motor or hydraulic pump \u2014 cannot operate. On electric tables the override is typically a fold-out crank that engages a socket on the table column. On hydraulic tables it is a manual release valve that vents oil from the lift cylinder. Both exist for one purpose: get the patient to a safe position when power is gone.<\/p>\n<p>The override is an emergency tool, not a positioning tool. Its job is to lower the table, level the tabletop, and in some designs restore Trendelenburg and side tilt so the patient can be transferred or the equipment repositioned. Using the crank for routine height changes wears the mechanism and wastes staff time. The hand control or foot pedal exists for normal operation; the crank exists for the failure you hope never happens.<\/p>\n<h2>Where the Emergency Crank Lives and How It Works<\/h2>\n<p><img decoding=\"async\" class=\"wp-image-2750\" src=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-quebec-1024w.webp\" alt=\"Operating table control pendant and column-mounted manual override panel with crank socket\" width=\"1024\" height=\"1365\" loading=\"lazy\" srcset=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-quebec-1024w.webp 1024w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-quebec-1024w-225x300.webp 225w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-quebec-1024w-768x1024.webp 768w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-quebec-1024w-9x12.webp 9w\" sizes=\"auto, (max-width: 1024px) 100vw, 1024px\" \/><\/p>\n<p>On most electric tables the crank socket sits on the column, below the tabletop, protected by a cap or door. The crank itself is stowed in a bracket on the column or base so it cannot wander off between uses. Operation is straightforward: remove the cap, insert the crank, and turn in the marked direction. One direction lowers the tabletop, the other raises it, and the label on the column states which is which. If the factory label is worn, facilities often mark the direction with paint or tape \u2014 a small step that removes guesswork during a real emergency.<\/p>\n<p>The crank engages the actuator drive train directly, so the tabletop can be raised or lowered by hand even with the motor completely dead. That direct mechanical path is why the crank remains the required fallback on tables where everything else is electric: it has no battery to deplete, no circuit to trip, and no software to fail.<\/p>\n<h2>Electric Tables: Crank, Battery, or Both<\/h2>\n<p><img decoding=\"async\" class=\"wp-image-2742\" src=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w.webp\" alt=\"Electric operating table with programmable hand control and manual override crank\" width=\"1536\" height=\"1152\" loading=\"lazy\" srcset=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w.webp 1536w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w-300x225.webp 300w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w-1024x768.webp 1024w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w-768x576.webp 768w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-electric-operating-table-product-photo-kilo-1536w-16x12.webp 16w\" sizes=\"auto, (max-width: 1536px) 100vw, 1536px\" \/><\/p>\n<p>Electric tables combine two fallbacks in different proportions. The battery backup keeps the electric drive alive through short outages \u2014 long enough to finish a critical step or complete a transfer. The manual crank covers everything else: a dead battery, a failed actuator, a tripped breaker that nobody can reset mid-case. The two systems work together, battery for the first minutes, crank for the long haul.<\/p>\n<p>Battery backup is a temporary power source with a finite, degrading service life, and it only works if someone tests it. The <a href=\"https:\/\/sanyangmedical.com\/operating-table-battery-backup-guide\/\">operating table battery backup guide<\/a> covers sizing, sulfation, and the test cadence that keeps the pack honest. For the crank, the equivalent discipline is a scheduled turn-and-verify during preventive maintenance.<\/p>\n<h2>Hydraulic Tables: Release Valves and Foot Pumps<\/h2>\n<p><img decoding=\"async\" class=\"wp-image-2826\" src=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w.webp\" alt=\"Mechanical hydraulic operating table with foot pump and manual cranks\" width=\"1920\" height=\"2560\" loading=\"lazy\" srcset=\"https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w.webp 1920w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w-225x300.webp 225w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w-768x1024.webp 768w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w-1152x1536.webp 1152w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w-1536x2048.webp 1536w, https:\/\/sanyangmedical.com\/wp-content\/uploads\/2026\/07\/sanyang-medical-mechanical-operating-table-product-photo-alpha-scaled-1920w-9x12.webp 9w\" sizes=\"auto, (max-width: 1920px) 100vw, 1920px\" \/><\/p>\n<p>Hydraulic operating tables fail differently from electric ones. A foot-pump table is already manual for height: the pedal builds pressure and a release valve lets the table down. Electro-hydraulic tables add a motorized pump, and their emergency control is a manual release valve \u2014 usually a lever or knob on the column \u2014 that vents oil from the cylinder so the tabletop lowers under controlled gravity.<\/p>\n<p>That release valve is the hydraulic equivalent of the electric crank, and it needs the same respect. A valve that has never been exercised can stick; a table lowered too fast through a cracked-open valve can alarm the whole room. Staff training should cover the correct opening sequence: crack the valve, confirm a controlled descent, then open fully only once the tabletop is near the target height.<\/p>\n<p>Which drive system suits which facility is a separate decision, and the <a href=\"https:\/\/sanyangmedical.com\/electric-vs-hydraulic-operating-table\/\">electric vs hydraulic operating table guide<\/a> compares failure modes, maintenance burden, and emergency behavior across both architectures.<\/p>\n<h2>What the Standards Require<\/h2>\n<p>IEC 60601-2-46, the particular standard for surgical operating tables, treats position-holding and emergency movement as essential performance. The general standard, <a href=\"https:\/\/webstore.iec.ch\/en\/publication\/67497\" target=\"_blank\" rel=\"noopener\">IEC 60601-1:2005+AMD1:2012+AMD2:2020<\/a>, defines the basic safety framework that applies to all medical electrical equipment. The particular standard adds the table-specific requirements, including the emergency lowering mechanism that must return the tabletop to a safe height during power loss, the maximum descent speed under load, and stability at maximum tilt.<\/p>\n<p>When a supplier&#8217;s test reports cite only the general standard, ask why the particular standard is missing. A table tested to IEC 60601-2-46 carries documented evidence for exactly the behaviors a manual override exists to protect. That report is the first document to request, and the <a href=\"https:\/\/sanyangmedical.com\/operating-table-height-adjustment-failure\/\">operating table height adjustment failure guide<\/a> explains how to separate drive faults from override faults when a table misbehaves in service.<\/p>\n<p>The same standard governs how fast the tabletop may descend when power is lost. Descent speed under full load matters because a table that drops quickly turns a controlled rescue into a patient-handling hazard, and a table that creeps down at a crawl keeps the team waiting mid-procedure. The test report should state the descent time from maximum height under rated load, not just the presence of the mechanism. Buyers who compare those numbers across models find real differences in emergency behavior that no brochure photograph will show.<\/p>\n<h2>Testing the Override Before It Is Needed<\/h2>\n<p>A crank that has never been turned can be seized, mislabeled, or missing. Facilities that run scheduled override tests as part of preventive maintenance catch those problems while the table is empty. The test takes minutes: cut the power or simulate a drive failure, lower and raise the tabletop through the override, verify the release valve vents smoothly, and record the result in the maintenance log.<\/p>\n<p>The buying decision should include the same demonstration. When you evaluate the <a href=\"https:\/\/sanyangmedical.com\/products\/operating-tables\/\">Sanyang Medical operating table range<\/a>, ask for the crank to be operated in front of you, check where the socket sits relative to anesthesia equipment, and confirm the release valve is reachable from both sides of the table. A model whose override is awkward to reach on paper will be impossible to use in practice.<\/p>\n<h2>Questions to Ask the Supplier Before Purchase<\/h2>\n<p>Four questions separate a table with a real override from a table with a checkbox:<\/p>\n<ul style=\"font-size:15px!important;\">\n<li><strong>Crank location<\/strong> \u2014 where does the socket sit, and can staff reach it with the table at full Trendelenburg?<\/li>\n<li><strong>Battery role<\/strong> \u2014 does the battery pack power emergency lowering, or only the crank?<\/li>\n<li><strong>Valve serviceability<\/strong> \u2014 if the release valve fails, is the mechanism serviceable on site or does the column need replacement?<\/li>\n<li><strong>Test evidence<\/strong> \u2014 is the override documented in the IEC 60601-2-46 test report, with descent speed and load conditions stated?<\/li>\n<\/ul>\n<p>Teams that understand the override before an incident handle it differently from teams that learn it under pressure. Run the demonstration at purchase, test the mechanism during maintenance, and keep the controls marked. The mechanical fallback is simple by design \u2014 keep it that way for the people who will use it once in a crisis.<\/p>\n<section class=\"sy-faq\">\n<h2>Operating Table Manual Override FAQ<\/h2>\n<div class=\"faq-card\">\n<h3>What is the manual override on an operating table?<\/h3>\n<p>The manual override is the mechanical fallback that moves or lowers the tabletop when the electric motor or hydraulic pump cannot operate. On electric tables it is an emergency crank; on hydraulic tables it is a manual release valve.<\/p>\n<\/div>\n<div class=\"faq-card\">\n<h3>Where is the emergency crank located on an operating table?<\/h3>\n<p>The crank socket sits on the table column below the tabletop, usually behind a cap or door, and the crank is stowed in a bracket on the column or base. The column label marks the lowering direction.<\/p>\n<\/div>\n<div class=\"faq-card\">\n<h3>Do hydraulic operating tables have a manual override?<\/h3>\n<p>Yes. Electro-hydraulic tables use a manual release valve that vents oil from the cylinder so the tabletop lowers under controlled gravity. Foot-pump tables are already operated manually for height adjustment.<\/p>\n<\/div>\n<div class=\"faq-card\">\n<h3>Is emergency lowering required by IEC 60601-2-46?<\/h3>\n<p>Yes. IEC 60601-2-46, the particular standard for surgical operating tables, requires an emergency lowering mechanism that returns the tabletop to a safe height during power loss, and the supplier&#8217;s test report should document it.<\/p>\n<\/div>\n<div class=\"faq-card\">\n<h3>How should the emergency crank be tested?<\/h3>\n<p>Facilities test the override during preventive maintenance: simulate a power failure, lower and raise the tabletop through the crank or release valve, and record the result.<\/p>\n<\/div>\n<\/section>\n<p><script type=\"application\/ld+json\">\n{\n  \"@context\": \"https:\/\/schema.org\",\n  \"@type\": \"FAQPage\",\n  \"mainEntity\": [\n    {\n      \"@type\": \"Question\",\n      \"name\": \"What is the manual override on an operating table?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"The manual override is the mechanical fallback that moves or lowers the tabletop when the electric motor or hydraulic pump cannot operate. On electric tables it is an emergency crank; on hydraulic tables it is a manual release valve.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"Where is the emergency crank located on an operating table?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"The crank socket sits on the table column below the tabletop, usually behind a cap or door, and the crank is stowed in a bracket on the column or base. The column label marks the lowering direction.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"Do hydraulic operating tables have a manual override?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Yes. Electro-hydraulic tables use a manual release valve that vents oil from the cylinder so the tabletop lowers under controlled gravity. Foot-pump tables are already operated manually for height adjustment.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"Is emergency lowering required by IEC 60601-2-46?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Yes. IEC 60601-2-46, the particular standard for surgical operating tables, requires an emergency lowering mechanism that returns the tabletop to a safe height during power loss, and the supplier's test report should document it.\"\n      }\n    },\n    {\n      \"@type\": \"Question\",\n      \"name\": \"How should the emergency crank be tested?\",\n      \"acceptedAnswer\": {\n        \"@type\": \"Answer\",\n        \"text\": \"Facilities test the override during preventive maintenance: simulate a power failure, lower and raise the tabletop through the crank or release valve, and record the result.\"\n      }\n    }\n  ]\n}\n<\/script><\/p>\n<p><script type=\"application\/ld+json\">\n{\n  \"@context\": \"https:\/\/schema.org\",\n  \"@type\": \"Article\",\n  \"name\": \"Operating Table Manual Override: Emergency Crank Operation Explained\",\n  \"description\": \"How the operating table manual override works: emergency crank on electric tables, release valves on hydraulic tables, IEC 60601-2-46 requirements, and the testing and procurement checks buyers should run.\",\n  \"author\": {\"@type\": \"Organization\", \"name\": \"Sanyang Medical\"},\n  \"publisher\": {\"@type\": \"Organization\", \"name\": \"Sanyang Medical\"},\n  \"datePublished\": \"2026-08-11\",\n  \"dateModified\": \"2026-08-11\",\n  \"mainEntityOfPage\": {\"@type\": \"WebPage\", \"@id\": \"https:\/\/sanyangmedical.com\/operating-table-manual-override-emergency-crank\/\"}\n}\n<\/script><\/p>\n<p><script type=\"application\/ld+json\" id=\"evo301-geo-ai-block\">\n{\n  \"definition\": \"Operating table manual override is the mechanical fallback system \u2014 an emergency crank on electric tables or a manual release valve on hydraulic tables \u2014 that lets staff move or lower the tabletop when the primary drive loses power. IEC 60601-2-46 requires an emergency lowering mechanism that returns the table to a safe height during power loss.\",\n  \"data_statements\": [\n    \"Electric operating tables use a fold-out crank that engages a socket on the table column, protected by a cap or door.\",\n    \"Electro-hydraulic operating tables use a manual release valve that vents oil from the lift cylinder for controlled lowering.\",\n    \"IEC 60601-2-46 treats position-holding and emergency movement as essential performance of surgical operating tables.\",\n    \"Battery backup powers the electric drive for short outages, while the crank remains the fallback for a dead battery or failed actuator.\",\n    \"Sanyang Medical's operating table range lists the override type and emergency lowering mechanism for each electric and electro-hydraulic model.\"\n  ],\n  \"qa_concise\": [\n    {\"q\": \"What is the operating table manual override?\", \"a\": \"The mechanical crank or release valve that moves the tabletop without electric power.\"},\n    {\"q\": \"Where is the emergency crank on a table?\", \"a\": \"In a socket on the table column, with the crank stowed in a bracket on the column or base.\"},\n    {\"q\": \"Do hydraulic tables have a manual override?\", \"a\": \"Yes, a manual release valve that vents oil from the lift cylinder.\"},\n    {\"q\": \"Is emergency lowering required by standard?\", \"a\": \"Yes, under IEC 60601-2-46 as part of essential performance.\"}\n  ]\n}\n<\/script><\/p>\n<p><script>\n(function(){var f=document.querySelectorAll('.yt-facade');for(var i=0;i<f.length;i++){(function(el){el.addEventListener('click',function(){var id=el.getAttribute('data-yt-id');el.innerHTML='<iframe style=\"position:absolute;top:0;left:0;width:100%;height:100%\" src=\"https:\/\/www.youtube.com\/embed\/'+id+'?autoplay=1\" frameborder=\"0\" allow=\"autoplay;encrypted-media\" allowfullscreen><\/iframe>';});})(f[i]);}})();\n<\/script><\/p>\n","protected":false},"excerpt":{"rendered":"<p>When power fails mid-procedure or a drive motor jams, the operating table manual override is the difference between a controlled repositioning and a stalled surgery. Every electric and electro-hydraulic surgical table carries a mechanical fallback \u2014 usually an emergency crank or a manual release valve \u2014 that lets staff move the tabletop without electricity. This &#8230; <a title=\"Operating Table Manual Override: Emergency Crank Operation Explained\" class=\"read-more\" href=\"https:\/\/sanyangmedical.com\/id\/operating-table-manual-override-emergency-crank\/\" aria-label=\"Baca selengkapnya tentang Operating Table Manual Override: Emergency Crank Operation Explained\">Baca Selengkapnya<\/a><\/p>","protected":false},"author":1,"featured_media":2780,"comment_status":"open","ping_status":"open","sticky":false,"template":"","format":"standard","meta":{"rank_math_title":"Operating Table Manual Override: Emergency Crank Operation | Sanyang","rank_math_description":"Operating table manual override explained: how the emergency crank and hydraulic release valve work, what IEC 60601-2-46 requires, and what buyers must test before purchase.","rank_math_focus_keyword":"operating table manual override","rank_math_robots":"","rank_math_canonical_url":"","rank_math_facebook_title":"","rank_math_facebook_description":"","rank_math_twitter_title":"","rank_math_twitter_description":"","_yoast_wpseo_title":"","_yoast_wpseo_metadesc":"","_yoast_wpseo_focuskw":"","_yoast_wpseo_canonical":"","_yoast_wpseo_meta-robots-noindex":"","_yoast_wpseo_meta-robots-nofollow":"","_yoast_wpseo_opengraph-title":"","_yoast_wpseo_opengraph-description":"","_yoast_wpseo_twitter-title":"","_yoast_wpseo_twitter-description":"","_aioseo_title":"","_aioseo_description":"","_aioseo_keywords":"","_aioseo_robots_default":"","_aioseo_robots_noindex":"","_aioseo_og_title":"","_aioseo_og_description":"","_aioseo_twitter_title":"","_aioseo_twitter_description":"","aiosp_title":"","aiosp_description":"","aiosp_keywords":"","_seopress_titles_title":"","_seopress_titles_desc":"","_seopress_analysis_target_kw":"","_seopress_robots_canonical":"","_seopress_robots_index":"","_seopress_robots_follow":"","_seopress_social_fb_title":"","_seopress_social_fb_desc":"","_seopress_social_twitter_title":"","_seopress_social_twitter_desc":"","_genesis_title":"","_genesis_description":"","_genesis_canonical":"","_genesis_noindex":"","_genesis_nofollow":"","slim_seo":"","footnotes":""},"categories":[108],"tags":[],"class_list":["post-3734","post","type-post","status-publish","format-standard","has-post-thumbnail","hentry","category-surgical-equipment-guides"],"_links":{"self":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/3734","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/comments?post=3734"}],"version-history":[{"count":1,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/3734\/revisions"}],"predecessor-version":[{"id":3893,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/posts\/3734\/revisions\/3893"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/media\/2780"}],"wp:attachment":[{"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/media?parent=3734"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/categories?post=3734"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/sanyangmedical.com\/id\/wp-json\/wp\/v2\/tags?post=3734"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}